IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
8 - 22 Section 8-4 - Reloading the software
8-4-11 Setup after Software loading
NOTE: Use this procedure if all partitions on the HDD have been erased during the software load.
NOTE: If only the software on C:\ has been replaced or updated, please go to: 8-4-12 "Verifications
after the software has been re-loaded" on page 8-23.
1.) Restore the Patient Archive and System Configurations from the backup you made before the
software loading.
- For instructions, please see “Data Backup and Restore” in the User Manual.
2.) With your recordings from before the software loading, available, continue with the setup
instructions starting in: Section 3-6 "Configuration" on page 3-24. Correct the settings as needed.
3.) Continue with: Section 3-8 "Connectivity setup" on page 3-38. Based on your recordings, correct
the settings as needed.
4.) Continue with: Section 3-9 "Options Setup" on page 3-52. Based on your recordings, correct the
settings as needed.
5.) After a software load, you should always calibrate the Front End, as described in: Section 6-6 "DC
Offset Calibration (Front End Alignment)" on page 6-11.
6.) Install the network printer (if any). Please refer to the Printer Driver Installation Manual.