Chapter 8 - Replacement procedures 8 - 91
8-6-7-2 Remove the Frame w/LCD and Touch Screen
1.) Place the Operator Panel, Upper, with the front down on a clean, antistatic surface.
2.) Remove the three (3x) fixing screws (Phillips #1) and the two screws at the D-SUB connector (3/16
inch nut driver), see: Figure 8-78.
NOTE: On some VIVID E9 / VIVID E7s, the openings for the USB connectors have EMC fingers
attached on the right and left sides of each opening, ref. Figure 8-79. Be careful to not loose
these EMC fingers when you remove the LCD Cover.
On newer VIVID E9 / VIVID E7s, the construction has been changed, so this issue has been
resolved.
3.) Lift (pull) the upper part of the LCD Cover so you can release it from the four protuding tabs.
Figure 8-78 Operator Panel, Upper, seen from the rear side
Figure 8-79 EMC fingers at openings for USB connectors
PULL TO REMOVE
PROTUDING TABS
(NOT VISIBLE IN THIS ILLUSTRATION)