GE RAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
8-52 Section 8-18 - Replacement of OPIO BUTTON SET
2.) Remove 5 probe holders. Refer to the figure below.
3.) Lift up main keyboard assy. Refer to the figure below.
Figure 8-65 Removing 5 probe holders
Figure 8-66 Lifting up main keyboard assy