8-94 Section 8-28 - Replacement of the NEST
2.) Remove the REAR CABINET BEZEL ASSY BT16. Refer to the figure below.
3.) Unscrew 7 screws and remove the REAR EMI COVER BRKT. Refer to the figure below.
Figure 8-140 Removing the Rear Cabinet Bezel Assy BT16
Figure 8-141 Removing the REAR EMI COVER BRKT
REAR CABINET BEZEL ASSY BT16