GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018)  BASIC SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5-71
3.) Recovering will task some time. 
Please wait until the final window (see figure below) appears and then press Enter. 
4.) In the displayed menus select [Exit] -> [Exit/Reboot]. 
5.) Remove the System DVD from the DVD/CD±R/RW Drive drive.
6.) Please wait until all processes are finished and the final “DOS” window (Figure 5-44 below) appear.
7.) Press A
 key to activate the Ultrasound Application Software [UIS].
Figure 5-42  Restore C; done 
Figure 5-43  Exit -> Exit/Reboot
The system boots with windows. An automated process was developed to check disk, assign an unique
computer name and matching drive letters. The system will reboot at least 2 times!
Figure 5-44  press “A” key to start UISApp