56 8555-9201-000 / 0109
Mechanical Separation / GEA Westfalia Separator
4.11 Alarm #13 - Fault series operation (option)
FAULT SERIES OPERATION
Line or
Pre-selection not ok
Alarm#13 ENTER
Text display Possible causes Action
“SERIES” key not pressed on both
controls.
Press “SERIES” key on both con-
trols.
Both separators not ready for opera-
tion
Check operational readiness of the
two separators.
Line (MPI) for series operation defec-
tive or missing.
Check line (MPI).
Switch position on MPI bus connector
not correct.
Switch position is at “OFF”.
Check switch position. Switch posi-
tion must be “ON”, terminator is ac-
tive.
Note :
MPI = Multi Point Interface
Help text information on the control system
ENT E R
Step 1 Acknowledge fault
HEL P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
FAULT SERIES OPERATION
Please check:
Series button pressed
On both devices
->
Page 2
Operational readiness
of both separators
MPI line
->
Page 3
Switch position on
MPI bus connector
b k ith ENTER
F1