8555-9201-000 / 0109 57
Mechanical Separation / GEA Westfalia Separator
4.12 Alarm #15 - WMS system fault
WMS – SYSTEM FAULT
Too much water
Separation “Stop”
Alarm#15 ENTER
Text display Possible causes Action
Water content in feed too high.
•
Check product feed.
Too much filling water.
T02 set too long.
• Check T02.
Filling water valve V2 leaks.
•
Check filling water valve V2.
Incorrect setting on evaluating relay.
• Check setting on evaluating relay.
T22 set too short.
•
Check setting T22.
• If necessary, extend T22 during this
phase.
Help text – information on the control system
ENT E R
Step 1 Acknowledge fault
HEL P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
WMS – SYSTEM FAULT
Please check:
Water content in
feed
->
Page 2
Filling water qty. T02
Valve V2
Setting on
evaluating relay
->
Page 3
Setting Z22
back with ENTER
F1