Refrigerated High-Speed Micro Centrifuge, 1730R
23
Error Possible Causes Actions
Error 9 RPM Sensor
· If the rotor recognition fails, this message comes up.
· The message will be cleared by coupling an appropriate rotor. Please refer to
[3.2 Rotor Coupling and Disassembling.] Disassemble and coupleacompatible
rotor and turn off and on the instrument to check out the status.
· If the error message does not disappear, please contact GENESPEED
®
Field
Service Engineer.
Error 11
Chamber
Temperature
· If the instrument is not reached to setting temperature within an hour, this
message is appeared.
· Please contact GENESPEED
®
Field Service Engineer.
Error 12 Temp. Sensor
· If there is a faulty in the temperature sensing of chamber or over heated, this
message is appeared.
· Please contact GENESPEED
®
Field Service Engineer.
Error 15
Motor Temperature
Sensor
· If the motor temperature sensor does not normally work, this message may
come up.
· Please contact GENESPEED
®
Field Service Engineer.
Error 16
Compressor
Temperature
· If the temperature of compressor is over heated up, this message is appeared.
· Please contact GENESPEED
®
Field Service Engineer.
Error 17
Data -
c o m m u n i c a t i o n
Error
· If the data can’t be transmitted fromthe main board to the display module, this
message is appeared.
· If the error message does not disappear, please contact GENESPEED
®
Field
Service Engineer.