33
C. Gerhardt GmbH & Co. KG
Either input or select a pre-entered program using
numeric keys.
Enter
Enter
7.
8.
11.3.1. Sample individual
1. Insert the inlet tubing into the digestion tube with the sulphuric digestion.
2. Press down quick clamping device and insert the digestion tube.
3. Check correct fit of the tube.
4. Close protection door and turn on equipment.
5. Press key "Analysis" in order to get into the analyse menu.
6.
Enter
Press "Sample individual".
Enter the normality of the titration acid used. If a
correction value has been determined enter it as
well. If a blank value has been defined in the cho-
sen program it is displayed. This value can be
overwritten.
Table:
0 = analysis without titration table
1 - 3 = analysis with titration table 1, 2 or 3 (see
chapter 7.2.3. Titration adjustment)