ULTIMA series ACCESS TO THE USER’S MENU
NU_ULTIMA_02_EN_1612 37
3. PUMP PRIMING
Selecting a program from main menu will open a menu in which
you can perform different functions related to program cycles.
Pressing the
icon will allow you to select and then prime a
chemical pump.
Select the desired chemical pump.
Once you press the
icon, the pump will run for 60
seconds or you can press
The
icon can be pressed at any time during the 60
seconds to stop the pump. After the pump stops, the chamber
will be filled with cold water for 30 seconds to help flush away
residual chemical and then the drain circuit will be energized for
30 seconds to send the solution to drain.