ULTIMA series MAINTENANCE OF THE WASHER
NU_ULTIMA_02_EN_1612 57
CHAPTER 8
MAINTENANCE OF THE WASHER
NEVER USE OR PLACE IN THE WASHER REMMANTS OF PRODUCTS SUCHS
AS ACETONE, SOLVANTS, OIL, SULFO-CROMIC, PETROL DERIVATIVES,
ACIDS (SULFURIC, NITRIC, CLORHIDRIC, EVEN AT LOW CONCENTRATIONS),
ETC.
DURING THE CLEANING OPERATIONS INSIDE THE CHAMBER, IT IS
CONVENIENT TO WEAR SAFETY GLOVES WITH GLOVES AND TO PAY
ATTENTION TO THE EVENTUAL EDGE TOOLS (NEEDLES, INSTRUMENTS…)
WHICH COULD BE PLACED INSIDE THE FILTERS OF THE BOTTOM OF THE
CHAMBER.
1. CHAMBER FILTERS
810LX – 815 LX – 820 LX – 910 LX – 1300 LX 1400 LX – 1400 LXP – 1600 LXP-
Clean the filters after each cycle.
Place them correctly.
Order of fitting: A -> B
1800 LXA