ULTIMA series MAINTENANCE OF THE WASHER
NU_ULTIMA_02_EN_1612 61
8. PREVENTATIVE MAINTENANCE SCHEDULE
FUNCTION
Daily (1)
Biannual (2)
1 Year (2)
2 Years (2)
5 Years (2)
Clean filter system in chamber.
Check chemical containers for any leakage.
Ensure chemical hoses are not pinched.
Clean exterior panels.
Inspect water supply hoses for cracks, bulges, and leaks.
Ensure the water hose seals and filters are clean and have no cracks.
Check chemical supply hoses for cracks, bulges and leaks.
Inspect internal tubing on chemical pumps for wear.
Check chemical level sensors for correct operation.
Ensure all panels are properly secured.
Check all internal hoses for cracks, bulges and leaks.
Ensure all hose clamps are properly tightened.
Check all column seals for leaks and cracks.
Check spray arm support seals for leaks and cracks.
Check spray arms and bushings for wear.
Check door seal and gasket for leakage.
Inspect the door springs, door wheels, door cable, hooks, gas spring, mounting hardware,
and door switch for proper operation.
X X X X
Change the Hepa filter of the dryer (if applicable)
Check recycling and emptying pump seals for leakage and quiet operation.
Check for lose electrical connections at components and electronic cards.
Verify the correct operation of all relays and their associated components.
Verify the correct operation of the fan of the dryer (if applicable)
Verify the correct operation of the non-return valve of the drying network
(if applicable)
X X X
(1) Daily maintenance must be handled by users staff.
(2) Others maintenances (bi-annual, 1 year, 2 years & 5 years) must be carry out by the GETINGE LANCER’s technical