ULTIMA series ALARMS
NU_ULTIMA_02_EN_1612 53
4. ALARMS TABLE
If an alarm occurs, the wash cycle is aborted and the appropriate
alarm message is shown in red.
Pressing the
icon will clear the alarm.
4.1 WASHER ALARMS LIST
1 DRAINING FAULT
During the draining phase, the washer as not
been drained within the configured drain
time.
- Check that drain pipe or drain
valve/pump is not obstructed.
- Verify operation of drain pump/valve.
- Check drain time.
9 LOW LEVEL ALARM
During the filling phase, the machine has not
achieved proper water level for pump
priming.
- Check the water supply, hoses, filter
screens, and valves.
10 FILLING FAULT : PURIFIED WATER
During the filling phase, the water high level
switch was not reached within the
configured filling time.
- Check opening of valves.
- Check pressure of water supply.
- Check filling time
30
ADDITIVE 1 INTAKE: OUT OF RANGE –
LOW
* optional ProKIT or validation monitor
During the cycle, the necessary amount of
product is not dosed.
- Check the tank level,
- Check the dosing pump, the flowmeter.
31
ADDITIVE 2 INTAKE: OUT OF RANGE –
LOW
* optional ProKIT or validation monitor
32
ADDITIVE 3 INTAKE: OUT OF RANGE –
LOW
* optional ProKIT or validation monitor
33
ADDITIVE 4 INTAKE: OUT OF RANGE –
LOW
* optional ProKIT or validation monitor
** optional extra alkaline dosing pump
86
ADDITIVE 5 INTAKE: OUT OF RANGE –
LOW
* optional ProKIT or validation monitor
** optional extra alkaline dosing pump