18 -
THE PROCESS
See Components in the TECHNICAL MANUAL for printer settings
and maintenance.
The system stores the last process so that an “emergency printout” can
be produced afterwards if there is a problem with transmission of the
printout or with the printer itself.
If the system is fitted with a SUPERVISOR, the printer is connected to
that. Otherwise it is normally connected to the control system. SUPER-
VISOR is an independent measuring system that records the process
data of the sterilizer. See also Control unit in the TECHNICAL MAN-
UAL for further information.
The illustration below is schematic and must not be used
as a template for evaluating individual process results.
13:47 29 06 04
SIGNATURE ………
..
6
PROCESS OK
5
00:24:58 71.2 132.5 85.2 0.995
PROCESS COMPLETE
00:23:39 75.5 132.5 85.2 0.039
EQUALIZATION
00:18:35 51.1 132.2 88.6 0.096
00:16:14 134.9 135.4 134.8 3.076
DRYING
HIGHEST TEMP 135.2C
LOWEST TEMP 134.9C
134.9C
00:16:08 135.0 135.5 134.9 3.094
4
POST-TREATMENT
00:12:07 134.1 134.3 134.0 3.039
STERILIZING
00:09:10 106.5 132.6 103.6 1.088