34 -
OPERATION
General advice when using the sterilizer
Keep the sterilizer door closed when the unit is not in use.
Read the section “General advice on packaging material” in the
guide below.
If it is difficult to get items dry, the post-treatment stage can be mod-
ified. See the guide “Cycle modification according to the type of
load” further down.
Read the manual for the medical device to be sterilized before start-
ing the process.
Be alert to everything that appears unusual such as leaks, humming
solenoid valves, stickily mechanical devices etc. Remedy the situa-
tion before it becomes a malfunction.
Heating of load
Heating of items to be processed may jeopardie sterilization result due
to poor humidification. Make sure appropriate process has been chosen
before loading the sterilizer.
TEMP. INTER-
LOCK
The independent safety interlocking of the
door function, as controlled by the load tem-
perature switch, or the Supervisor, is faulty.
NOTE: Not present on all sterilizers.
I/O FAULT The control system has lost communication
with the input or output cards.
Call a service engineer immediately.
WATER
SUPPLY
The water level in the tank is too low.
COMPR.AIR
SUPPLY
Pressure of incoming compressed air is too
low.
DEGAS. TANK
FAIL
The temperature in the degassing tank has
fallen below the permitted value.
NOTE: Not present on all sterilizers.