OPERATION INSTRUCTIONS
Daily preparations
Learn the functions of the control buttons and signal lamps by stud-
ying the Instruments and Control unit chapters. Find out how the
user-programmable parameters are set on this unit.
Open valves for water, air and, where applicable, steam.
Switch on the mains switch of the unit.
Check that recording instruments and printers have sufficient paper.
The paper of recording instrument shows a red line when it is nearly
finished.
Run the first steam sterilization process of the day with the chamber
empty. This heats up and drains the system pipework.
Running a process
Select the desired program. The selected program is indicated on the
control panel.
Parameter password
Before a selectable parameter can be changed, a parameter password
must be entered, after which all parameters (see Program combina-
tion) can be changed.
Password-protected programs
Programs that are not intended to be used routinely are protected
with a password. Some degree of consideration is called for before a
password-protected programs is chosen. Examples include test pro-
grams, programs for emergency situations, programs with selectable
parameters or programs using some sterilization medium other than
steam.
Load the sterilizer.
Start via operator panel Avanti
Make sure that cycle media are availa-
ble.
Close the door by pressing
Close
.
If the “Door locked” indicator is lit and
the
Start
button can be selected, the
machine is ready to start.