P 0 1 Wrapped goods, textiles, porous load
RAX2000
Parameter Range Delivered Actual
Negative pulses - - - 3 3 ................
Positive pulses - - - 5 5 ................
1 Sterilizing temperature °C 134 134 ................
2 Sterilizing time min 3 - 7 4 ................
3 Postvacuum time min 0 - 90 5 ................
4 Postpuls steam - - - 0 - 25 0 ................
5 Postpuls air - - - 0 - 25 0 ................
P 0 2 Wrapped, heat sensitive material, rubber, plastic, porous load
RAX2001
Parameter Range Delivered Actual
Negative pulses - - - 3 3 ................
Positive pulses - - - 5 5 ................
1 Sterilizing temperature °C 121 121 ................
2 Sterilizing time min 16 - 20 16 ................
3 Postvacuum time min 0 - 90 5 ................
4 Postpuls steam - - - 0 - 25 0 ................
5 Postpuls air - - - 0 - 25 0 ................
P 0 3 Rapid process for single, open instrument
RAX2002
Password required!
Parameter Range Delivered Actual
Negative pulses - - - 1 1 ................
Positive pulses - - - 2 2 ................
1 Sterilizing temperature °C 134 134 ................
2 Sterilizing time min 3 - 90 4 ................
3 Postvacuum time min 0 - 90 3 ................
B3111_Rev 2
Program combination 2530290-090
Liquids must not be processed with any of these programs.
Hazardous waste and explosive materials must not be processed in this sterilizer.
Door blocking key
The door does not close when the key is in the locked position and the door is in open
position, as indicated by the symbol. When entering or cleaning the chamber the key
must be removed from the locking device.