P 0 4 Bowie & Dick test
RAX2003
Password required!
Parameter Range Delivered Actual
Negative pulses - - - 3 3 ................
Positive pulses - - - 5 5 ................
1 Sterilizing temperature °C 121 - 135 134 ................
2 Sterilizing time min 0 - 15 1 ................
3 Postvacuum time min 0 - 90 3 ................
P 05 Automatic leak rate test
RAX2004
Password required!
Parameter Range Delivered Actual
1 Stabilizing time min 10 - 90 10 ................
2 Test time min 10 10 ................
B3111_Rev 2