OPERATION -
25
OPERATION
Program start
Where applicable, the operator must state process parameters and other
information requested by the control system. On equipment where an
authorisation code for programs not intended for routine sterilization is
activated, the operator must also enter an authorisation code immedi-
ately after startup.
Via operator panel Avanti
To start a cycle:
Process startup is prevented if:
there is an unacknowledged alarm
the door is not closed and interlocked in accordance with the condi-
tions
the control system has detected an interlocking error (affects door
safety)
the keyswitch for manual stepping is activated
the keyswitch for door blocking has been activated (only if there is a
keyswitch)
media supply, eg steam or cooling water, is not available (only if the
sterilizer is equipped with sensors for this).
the jacket temperature has not been reached or is too high (only if the
process includes jacket heating)
Select the desired cycle.
Load goods and close the door.
Wait until the Door closed indicator lights
up and the
Start
button becomes visible.
Now the machine is ready to start.
Start the cycle.