OPERATION -
33
LOW TEMPER-
ATURE
The chamber temperature has been below the
permitted limit during the sterilization phase.
PHASE TIME-
OUT
The phase named on the display exceeded the
time limits for completion. Check whether the
cause is a component fault or media fault.
POST. TIMEOUT The post-treatment exceeded the times limits
for completion. Check whether the cause is a
component fault or media fault.
MAINTENANCE This message appears when the programmed
service interval has elapsed and the current
process is complete. The message continues to
be displayed until it is de-activated by the
service technician.
The autoclave can still be used.
Call a service engineer immediately.
MANUAL OUT-
PUTS
A digital output has been manually switched
off or on. Call a service technician to reset the
output to auto mode.
LEAK RATE
FAIL
An automatic leak test has failed.
SUPERVISOR
ERROR
The Supervisor’s independent monitoring of
process time and temperature conditions has
indicated that these have not been fulfilled and
have triggered an alarm.
POWER
FAILURE
There has been a power failure longer than 10
seconds.
EMERGENCY
STOP
The program has been stopped with the emer-
gency stop button.
STEAM GENER-
ATOR
The supply of feedwater to the steam genera-
tor has stopped working.
DOOR INTER-
LOCK
The independent safety interlocking of media
admission to the chamber, as controlled by the
door switch, is faulty.
GASKET INTER-
LOCK
The independent safety interlocking of media
admission to the chamber, as controlled by the
gasket pressure switch, is faulty.
PRESS. INTER-
LOCK
The independent safety interlocking of the
door function, as controlled by the chamber
pressure switch,pressure monitor or the
Supervisor, is faulty.
Error code Explanation