48 -
OPERATION INSTRUCTIONS
Measures in case of alarm
Programs that are installed to be ended automatically
•
The program ends automatically.
An emergency stopped process or program that did not finish auto-
matically.
•
Note the process phase during which the fault occurred.
Options for alterative actions are described in the chapter “Alarm” in
chapter Operation.
After a fault in a process, the password must be entered before the load-
ing side door can be opened.
On power failure or if the control switch or working switch
of the sterilizer had to be turned off during a process, there
is a risk of leakage at the door. Safety measure when there
is positive pressure in the chamber:
KEEP WELL AWAY FROM THE DOOR AND CALL A TECH-
NICIAN!
ALWAYS LEAVE THE PREMISES and call a technician if
the chamber pressure is in the RED area of the pressure
gauge.
•
Silence the alarm signal by pressing
Silence alarm
signal
.