6.2.2.2 From the wall-mounted control keypad (on VCS only)
1
2
3
Fig.56: Battery tests from single the wall-mounted keypad
Running a battery backup test
1. Turn off the light.
2. Press Battery backup test 1 .
Ø If the test is successful, the battery level indicator 2 flashes green. If the test fails, the
battery level indicator 2 flashes red.
3. If the test fails, contact the Getinge technical service department.
4. Press Battery backup test 1 again and hold until the button turns off.
Ø The light remains on at Level 3 and the system is ready for use.
Running a battery life test (only with a Getinge backup)
1. Turn off the light.
2. Press Battery lifetime test 3 and hold until the button is backlit.
Ø If the test is successful, the battery level indicator 2 flashes green. If the test fails, the
battery level indicator 2 flashes red.
3. If the test fails, contact the Getinge technical service department.
Ø The light turns off when the test is complete.
4. Press Battery lifetime test 3 again and hold until the button turns off.
NOTE
The battery lifetime test can be stopped at any time by pressing Battery lifetime
test 3 .