6.2.3 Electrical safety tests
NOTE
Electrical safety measurements must be carried out using an IEC62353-compat-
ible electrical safety tester. The earth resistance should be less than or equal to
300mΩ.
Fig.58: Electrical safety test
The measurement should be made between the flange 1 and the edge of the connector located
in the centre of the lighthead 2 .