Safety instructions
8 / 56
Maquet Volista
NM 01780 EN 03
2 Safety instructions
WARNING!
Risk of electric shock
Anyone not trained in installation, maintenance or decommissioning opera-
tions is exposed to the risk of injury or electric shock.
Installation, maintenance and decommissioning of the device or components
of the device must be performed by a Getinge technician or a Getinge-trained
service technician.
WARNING!
Risk of infection
A servicing or cleaning operation may result in contamination of the surgical
site.
Do not perform servicing or cleaning operations when the patient is present.
WARNING!
Risk of infection
If no decontamination is carried out on the device before servicing work,
there is a risk of infection for anyone handling the device or any of its com-
ponents.
Make sure that the device is fully decontaminated before any servicing work
is conducted.
WARNING!
Risk of burns
During maintenance operations, certain accessible parts may be hot immedi-
ately after use of the device.
Allow the device to cool down before performing any service.
WARNING!
Risk of injury
Parts (e.g. tools, screws and colts) are liable to fall during a technical proced-
ure.
Please limit your presence under the device during technical procedures,
and use the appropriate safety equipment (e.g. safety hat, safety glasses).
WARNING!
Risk of infection
Lightweight parts from the device may fall onto the surgical site.
Check that all fastenings, covers, cover plates and bumpers on the device
are properly in place.
WARNING!
Risk of electric shock or injury
The use of screws or spare parts other than those supplied by the manufac-
turer may damage the device.
Use only screws and spare parts supplied by the manufacturer.