RecordCurrentSequenceNumber
If recording is running, this feature returns the number of the currently active
sequence.
Access Read
Interface IInteger
Unit --
Range 0 …16
Invalidation --
Precondition --
Note --
RecordBurstCnt
This feature defines the number of consecutive frames to record after an
external trigger or an image trigger occurs. Only available in single recording
mode.
Access Read / write
Interface IInteger
Unit --
Range 0 …1022, 0 means to record until the buffer is filled
Invalidation --
Precondition CameraState = Idle
RecordMode = Single
RecordWhileTriggerActive = False
Note Each external trigger causes the camera to record
RecordBurstCnt frames. The frames are attached to
previously sampled frames already stored in the cam-
era buffer. New frames/frame sequences can be
added, until the buffer is full.
only for single recording. If RecordMode = Continuous,
the feature will be locked and the value will be invalid.
Feature description
MikrotronGmbH / mini1 / CAMMC1370/71-RG / 2021 36 - 67