3-35
CHAPTER 3 Settings and Measurement
4ch Voltage / Temperature Terminal (GS-4VT: optional) setting
Performs the setting related to the input of 4ch voltage / temperature terminal.
The Setting screen is displayed at the end of the group by connecting the 4ch voltage / temperature terminal
and turning On the GL840’s power.
(1)-22 Input setting
This is used to select the input conditions.
Setting Description
1ch to 4ch Off The input signal is not measured. The waveform and digital display also are not performed.
DC Used to measure the direct-current voltage.
TEMP Used to measure the temperature.
5ch to 8ch No AMP Because the amplier cannot be installed, the “No AMP” is displayed.
(1)-23 Sensor setting
This is used to select the sensor type to be used.
Setting Description
TC-K Set when the thermocouple to be connected is K-type.
TC-T Set when the thermocouple to be connected is T-type
(1)-24 Range setting
Setting Description
TEMP 200C xed range
DC 20, 50, 100, 200, 500mV; 1, 2, 5, 10, 20, 50V; 1-5V
(1)-25 EU setting
This is used to unit-convert the measurement signal.
Refer to “EU Function, Meas. Value, EU Output Value, and Reads the current temperature measurement
value” in “(1)-5 EU (Scaling settings)” in “3.4 Setting Menus”. In the case of Voltage, refer to “<Setting
example: For voltage input>”.
(1)-26 Other setting
Annotation text, Inter-CH calculation, span, line width, trace, and zero point, etc. can be set.
Refer to “(1)-6 Misc.” in “3.4 Setting Menus”.