4-6
CHAPTER 4 Specication
GL840-WV: B-565
Item Description
Number of input channels 20ch (200ch available when used with the extension terminal base)
* Possible to direct-connect or connect with the extension terminal connection cable
(sold separately) between the GL840 and terminal base, or between terminal
bases.
Input terminal type M3 screw type terminals (Rectangular at washer)
Input method Photo MOS relay scanning system
All channels isolated, balanced input
* Terminal b to be used to connect the Resistance bulb is shorted within all
channels.
Scan speed 10 ms/1 ch maximum
Measurement ranges Voltage: 20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50, 100 V; 1-5 V F.S.
Temperature
Thermocouples : K, J, E, T, R, S, B, N, C (WRe5-26)
Resistance bulb : Pt100, JPt100, Pt1000 (IEC751)
Temperature range: 100°C, 500°C, 2000°C
(In the case of Fahrenheit: 150°F, 750°F, 3000°F)
Humidity: 0 to 100% (voltage 0 V to 1 V scaling conversion) xed
Measurement accuracy
(23°C ±5°C)
• When 30 minutes or more have
elapsed after power was switched on
• Sampling 1 s/10 ch
• Filter ON (10)
• GND connected
Voltage: ± (0.05% of F.S. + 10μV)
Temperature
•
Thermocouple
Thermocouple
Measurement Temperature
Range (°C)
Measurement Accuracy
R/S 0 ≤ TS ≤ 100℃
100 < TS ≤ 300℃
R : 300 < TS ≤ 1600℃
S : 300 < TS ≤ 1760℃
±4.5℃
±3.0℃
±2.2℃
±2.2℃
B 400 ≤ TS ≤ 600℃
600 < TS ≤ 1820℃
±3.5℃
±2.5℃
K -200 ≤ TS ≤ -100℃
-100 < TS ≤ 1370℃
±1.5℃
±0.8℃
E -200 ≤ TS ≤ -100℃
-100 < TS ≤ 800℃
±1.0℃
±0.8℃
T -200 ≤ TS ≤ -100℃
-100 < TS ≤ 400℃
±1.5℃
±0.6℃
J -200 ≤ TS ≤ -100℃
-100 < TS ≤ 100℃
100 < TS ≤ 1100℃
±1.0℃
±0.8℃
±0.6℃
N -200 ≤ TS < 0℃
0 ≤ TS ≤ 1300℃
±2.2℃
±1.0℃
C (W) 0 ≤ TS ≤ 2000℃ ±1.8℃
Reference contact compensation accuracy ±0.3℃
* Thermocouple diameters T - K: 0.32 φ, others: 0.65 φ
•
Resistance bulb
Type
Measurement Temperature
Range (°C)
Applied
current
Measurement
Pt100 -200 ≤ TS ≤ 100℃ 1mA ±0.6℃
100 < TS ≤ 500℃ ±0.8℃
500 < TS ≤ 850℃ ±1.0℃
JPt100 -200 ≤ TS ≤ 100℃ 1mA ±0.6℃
100 < TS ≤ 500℃ ±0.8℃
Pt1000 -200 ≤ TS ≤ 100℃ 0.3mA ±0.6℃
100 < TS ≤ 500℃ ±0.8℃
* 3-wire system