3-21
CHAPTER 3 Settings and Measurement
3.4 Setting Menus
When you press the [MENU] key during Free Running, the following menu screens appear.
WLAN ※1 OTHERI/FTRIGAMP DATA
* When optional wireless unit is not installed, these menu screen are not displayed.
(1) AMP settings
This menu is used to specify input signal-related settings.
<Analog settings> <Logic/Pulse settings>
Setting Selections available
Input Standard terminal, Withstand high-
voltage high-precision terminal,
Screwless terminal
Off, DC, TEMP, RH
GS sensor and terminal/addapter
(optional)
Temperature and humidity sensor (GS-TH): Off, TEMP, DEW, ACC.
TEMP
3-axis acceleration / temperature sensors (GS-3AT): Off, TEMP /
ACC. X / ACC. Y / ACC. Z
4ch voltage / temperature terminal (GS-4VT): Off, DC, TEMP
4ch thermistor (GS-4TSR): Off, TEMP
Illumination / ultraviolet sensor (GS-LXUV): Off, ILLUM. / UV / ACC.
ILLUM. / ACC. UV
CO2 sensor (GS-CO2): Off, CO2
Adapter for AC current sensor (GS-DPA-AC): Off, CURRENT /
POWER, ACC. POWER
Standard terminal
and Withstand
high-voltage high-
precision terminal
Range [Voltage] 20, 50, 100, 200, 500mV; 1, 2, 5, 10, 20, 50, 100V; 1-5V
[Temperature] Sensor Thermocouple: TC-K, TC-J, TC-T, TC-R, TC-E, TC-B, TC-S, TC-N,
TC-C
Resistance bulb: Pt100, JPt100, Pt1000
Range 100, 500, 2000°C
[Humidity] 100% xed
Filter Off, 2, 5, 10, 20, 40
Temperature and
humidity sensor
(GS-TH)
Range [Temperature] / [Dew-point
temperature] / [Accumulated
temperature]
100°C xed
[Humidity] 100% xed
Filter None
3-axis acceleration
/ temperature
sensor (GS-3AT)
Range [Acceleration] Sensor Peak / Ave. / RMS
Range 2, 5, 10G; 20, 50, 100m/s
2
[Temperature] 100°C xed
Filter None
4ch voltage /
temperature
terminal (GS-4VT)
Range [Voltage] 20, 50, 100, 200, 500mV; 1, 2, 5, 10, 20, 50V; 1-5V
[Temperature] Sensor Thermocouple: TC-K , TC-T
Range 2000°C xed
Filter None
4ch thermistor
(GS-4TSR)
Range [Temperature] Sensor Thermistor: TSR-A, TSR-J
Range 200°C xed
Filter None