1-4
CHAPTER 1 General Description
(7) Pollution degree
• POLLUTION DEGREE 2 in accordance with IEC60664
•If condensation occurs...
Condensation occurs in the form of water droplets on the device surfaces and interior when the GL840
is moved from a cold to a warm location. Using the GL840 with condensation will cause malfunctioning.
Wait until the condensation has disappeared before turning on the power.
Warming-up Before Use
The GL840 should be allowed to warm up with the power turned on for approximately 30 minutes to ensure
that it operates according to the specied performance.
Conguration When in Use
Do not use the GL840 standing upright or at an angle. It must always be laid at or inclined on the stands.
<Usage Conguration>
Do not block the air vent on the GL840, as this will cause malfunctioning.
Measurement accuracy may not be satisfactory if the system is used in a condition other than described
above.