4-4
CHAPTER 4 Specication
GL840-M: B-564 / B-564SL
Item Description
Number of input
channels
Standard terminal 20 channels (Up to 200 channels when using with the expansion terminal base)
* Between GL840 and terminal and between terminals can be direct-connected or
with an expansion terminal connection cable (sold separately)
Screwless terminal 20 channels (Up to 200 channels when using with the expansion terminal base)
* Between GL840 and terminal and between terminals can be direct-connected or
with an expansion terminal connection cable (sold separately)
Input terminal
type
Standard terminal M3 screw type terminals (Rectangular at washer)
Screwless terminal Screwless terminal
Input method Photo MOS relay scanning system
All channels isolated, balanced input
* Terminal b to be used to connect the resistance bulb is shorted within all channels.
Scan speed 10 ms/1 ch maximum
Measurement ranges Voltage: 20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50, 100 V; 1-5 V F.S.
Temperature
Thermocouples : K, J, E, T, R, S, B, N, C (WRe5-26)
Resistance bulb : Pt100, JPt100, Pt1000 (IEC751)
Temperature range : 100°C, 500°C, 2000°C
(In the case of Fahrenheit: 150°F, 750°F, 3000°F)
Humidity: 0 to 100% (voltage 0 to 1 V scaling conversion) xed
Measurement accuracy
(23°C ±5°C)
• When 30 minutes or more have
elapsed after power was switched on
• Sampling 1 s/20 ch
• Filter ON (10)
• GND connected
Voltage: 0.1% of F.S.
Temperature
•
Thermocouple
Thermocouple
Measurement Temperature
Range (°C)
Measurement Accuracy
R/S 0 ≤ TS ≤ 100℃
100 < TS ≤ 300℃
R : 300 < TS ≤ 1600℃
S : 300 < TS ≤ 1760℃
±5.2℃
±3.0℃
± (0.05% of rdg +2.0℃)
± (0.05% of rdg +2.0℃)
B 400 ≤ TS ≤ 600℃
600 < TS ≤ 1820℃
±3.5℃
± (0.05% of rdg +2.0℃)
K
-
200 ≤ TS ≤
-
100℃
-100 < TS ≤ 1370℃
± (0.05% of rdg +2.0℃)
± (0.05% of rdg +1.0℃)
E
-
200 ≤ TS ≤
-
100℃
-
100 < TS ≤ 800℃
± (0.05% of rdg +2.0℃)
± (0.05% of rdg +1.0℃)
T
-
200 ≤ TS ≤
-
100℃
-
100 < TS ≤ 400℃
± (0.1% of rdg +1.5℃)
± (0.1% of rdg +0.5℃)
J
-
200 ≤ TS ≤
-
100℃
-
100 < TS ≤ 100℃
100 < TS ≤ 1100℃
±2.7℃
±1.7℃
± (0.05% of rdg +1.0℃)
N
-
200 ≤ TS < 0℃
0 ≤ TS ≤ 1300℃
± (0.1% of rdg +2.0℃)
± (0.1% of rdg +1.0℃)
C (W) 0 ≤ TS ≤ 2000℃ ± (0.1% of rdg +1.5℃)
Reference contact compensation accuracy ±0.5℃
* Thermocouple diameters T - K: 0.32 φ, others: 0.65 φ
•
Resistance bulb
Type
Measurement Temperature
Range (°C)
Applied
current
Measurement
Pt100
-
200 to 850℃ (FS=1050℃) 1mA ±1.0℃
JPt100
-
200 to 500℃ (FS=700℃) 1mA ±0.8℃
Pt1000
-
200 to 500℃ (FS=700℃) 0.3mA ±0.8℃
* 3-wire system