2-26
CHAPTER 2 Checks and Preparation
2.14
Mounting the Extension Terminal Base (Optional) and Extension Terminal Cable (Optional)
2.14.1 Mounting the extension terminal base
This section describes how to mount the extension terminal base.
When mounting the extension terminal base on the GL840, please make sure that the GL840’s power is
turned OFF.
Prepare the sold separately extension terminal base and the extension terminal connection cable.
B-566 Extension Terminal Base
Extension terminal base: 1 unit Connection plate: 1 pc. M4 x 6 at-head screw: 4 pcs.
B-567 Extension terminal connection cable (Select from two types of cables)
Extension terminal connection cable
(50 cm: B-567-05): 1 pc.
Extension terminal connection cable
(2 m: B-567-20): 1 pc.
Mounting
(1) Remove the Standard terminal or the Withstand high-voltage high-precision terminal mounted on the
GL840.
(2) Insert the lock tabs at the top of the terminal into the slots of the extension terminal base, and press the
terminal until the lock tabs at the bottom of the terminal are securely locked.
Insert
Press
* When installing the Withstand high-
voltage high-precision terminal (B-
565), please push it while pulling to
the front.