3-37
CHAPTER 3 Settings and Measurement
(1)-29 Sensor setting
This is used to select the sensor type to be used.
Setting Description
TSR-A Set when the thermistor is GS-103AT-4P.
TSR-J Set when the thermistor is GS-103JT-4P.
(1)-30 Range setting
The temperature range is xed at 200°C.
(1)-31 EU setting
This is used to unit-convert the measurement signal.
Refer to “EU Function, Meas. Value, EU Output Value, and Reads the current temperature measurement
value” in “(1)-5 EU (Scaling settings)” in “3.4 Setting Menus”.
(1)-32 Other settings
Annotation text, Inter-CH calculation, span, line width, and trace, etc. can be set.
Refer to “(1)-6 Misc.” in “3.4 Setting Menus”.
(1)-33 Logic / Pulse setting
This is used to perform the setting related to the digital input to the GL840.
Refer to “(1)-7 Logic/Pulse” in “3.4 Setting Menus”.
However, for Group settings, LOGIC5 to LOGIC8 and PULSE5 to PULSE8 are displayed.
When the pulse is measured by the GL100-WL or GS sensor, the pulse measurement is executed in GL100-
WL or GS sensor side.
The fastest sampling speed of the GL100-WL and GS sensor is 500 ms.
When the sampling speed of the GL840 is faster than 500 ms, there is a possibility that the error between
the value captured in 500 ms by the GL100-WL and GS sensor and the pulse captured by the GL840 may
occur, in order to execute the calculation processing within the sampling interval of GL840.