EasyManua.ls Logo

GRAPHTEC midi LOGGER GL240 - Input Unit Specifications

GRAPHTEC midi LOGGER GL240
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-4
CHAPTER 4 Specication
Input Unit Specications
Item Description
Number of input channels 10 channels
Input terminal type M3 screw type terminals (Rectangular at washer)
Input method Photo MOS relay scanning system
All channels isolated, balanced input
Scan speed 10 ms/1 ch maximum
Measurement ranges Voltage: 20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50, 100 V; 1-5 V F.S.
Temperature
Thermocouples : K, J, E, T, R, S, B, N, W (WRe5-26)
Humidity: 0 to 100% (voltage 0 to 1 V scaling conversion) * Use the B-530 (optional)
Measurement accuracy
(23°C ±5°C)
• When 30 minutes or more have
elapsed after power was switched on
• Sampling 1 s/10 ch
• Filter ON (10)
• GND connected
Voltage: 0.1% of F.S.
Temperature
• Thermocouple
Thermocouple
Measurement Temperature
Range (°C)
Measurement Accuracy
R/S 0 TS ≤ 100℃
100 < TS 300℃
R : 300 < TS 1600
S : 300 < TS 1760℃
±5.2℃
±3.0℃
± (0.05% of rdg +2.0℃)
± (0.05% of rdg +2.0℃)
B 400 TS ≤ 600
600 < TS ≤ 1820℃
±3.5℃
± (0.05% of rdg +2.0℃)
K
-
200 TS ≤
-
100
-100 < TS ≤ 1370℃
± (0.05% of rdg +2.0℃)
± (0.05% of rdg +1.0℃)
E
-
200 TS
-
100
-
100 < TS ≤ 800℃
± (0.05% of rdg +2.0℃)
± (0.05% of rdg +1.0℃)
T
-
200 TS
-
100
-
100 < TS ≤ 400℃
± (0.1% o f rdg +1.5℃)
± (0.1% o f rdg +0.5℃)
J
-
200 TS
-
100
-
100 < TS ≤ 100℃
100 < TS ≤ 1100℃
±2.7
±1.7
± (0.05% of rdg +1.0℃)
N
-
200 TS < 0℃
0 TS ≤ 1300℃
± (0.1% o f rdg +2.0℃)
± (0.1% o f rdg +1.0℃)
W 0 TS 2000℃ ± (0.1% o f rdg +1.5℃)
Reference contact compensation accuracy ±0.5℃
* Thermocouple diameters T, K: 0.32 φ, others: 0.65 φ
Reference contact compensation
accuracy
Internal/External switching
A/D converter Method : ΔΣ method
Resolution :16-bit (Effective resolution: About 1/40000 of the +/- range)
Temperature coefcient Gain : 0.01% of F.S./°C
Zero : 0.02% of F.S./°C
* Zero occurs at the sampling of 10, 20, and 50 ms.
Input resistance 1MΩ ±5%
Allowable signal source resistance Within 30
Maximum permissible input voltage Between +/– terminals : 20mV to 1V range (60Vp-p)
2V to 100V range (110Vp-p)
Between input terminal/input terminal : 60 Vp-p
Between input terminal/GND : 60 Vp-p
Withstand voltage Between input terminal/input terminal : 350 Vp-p 1 minute
Between input terminal/GND : 350 Vp-p 1 minute
Insulation resistance Between input terminal/GND : 50MΩ or more (at 500 VDC)
Common mode rejection ratio 90 dB or more (50/60 Hz; signal source 300Ω or less)
Noise 48 dB or more (with +/– terminals shorted)
Filter Off, 2, 5, 10, 20, 40
Filter operation is on a moving average basis.
The average value of the number of set samples is used.
If the sample interval exceeds 5 seconds, the average value of data obtained in a
sub-sample (5 seconds) is used.