3-19
CHAPTER 3 Settings and Measurement
3.4 Setting Menus
When you press the [MENU] key during Free Running, the following menu screens appear.
<When optional wireless unit is installed>
<When optional wireless unit is not installed>
LAN OTHI/FTRGAMP DAT
OTHRI/FTRIGAMP DATA
(1) AMP settings
This menu is used to specify input signal-related settings.
<Analog settings> <Logic Pulse settings>
The GL100-WL+GS sensor shown below is required to install the optional wireless unit.
For more information about the GL240 and GL100-WL wireless connection, refer to "(5) LAN setting” and
the operation procedure example in "3.4 Setting Menus”.
Setting Selections available
Input Range [Voltage] 20, 50, 100, 200, 500mV; 1, 2, 5, 10, 20, 50, 100V; 1-5V
[Temperature] Sensor Thermocouple: TC-K, TC-J, TC-T, TC-R, TC-E, TC-B, TC-S, TC-N,
TC-W
Range 2000°C xed
[Humidity] 100% xed
Filter Off, 2, 5, 10, 20, 40
Temperature and
humidity sensor
(GL100-WL+GS-
TH)
Range [Temperature] / [Dew-point
temperature] / [Accumulated
temperature]
100°C xed
[Humidity] 100% xed
Filter None
3-axis acceleration
/ temperature
sensor (GL100-
WL+GS-3AT)
Range [Acceleration] Sensor Peak / Ave. / RMS
Range 2, 5, 10G; 20, 50, 100m/s
2
[Humidity] 100% xed
Filter None
4ch voltage /
temperature
terminal (GL100-
WL+GS-4VT)
Range [Voltage] 20, 50, 100, 200, 500mV; 1, 2, 5, 10, 20, 50V; 1-5V
[Humidity] Sensor Thermocouple: TC-K , TC-T
Range 2000°C xed
Filter None
4ch thermistor
(GL100-WL+GS-
4TSR)
Range [Humidity] Sensor Thermistor: TSR-A, TSR-J
Range 200°C xed
Filter None