3-43
CHAPTER 3 Settings and Measurement
(2) DATA settings
This is used to specify capture-related items and calculations.
Setting Selections available
Sampling 10, 20, 50, 100, 125, 200, 250, 500ms, 1, 2, 5, 10, 20, 30s, 1, 2, 5, 10, 20, 30min,
1h; External
* Available sampling intervals vary depending on the input settings and the number of
channels to be used.
* Refer to "(2)-1 Sampling Interval" for details.
Capture destination SD1 (Internal memory), SD2 (SD memory card)
File Name * See the description of a captured data le name given below.
Ring/Relay capture Off, Ring, Relay
Ring capture Number of recording points
AC Line Filter Off, On
Backup Backup Interval Off, 1, 2, 6, 12, 24 hours
Backup Destination SD1 (Internal memory), SD2 (SD memory card), FTP
Save Folder Folder name
Calc. Settings 1 Off, Average, Max, Min, Peak, RMS
Calc. Settings 2 Off, Average, Max, Min, Peak, RMS
Captured data le name
Setting Selections available
Folder (File) Capture destination : SD1 (Internal memory), SD2 (SD memory card)
Folder : Text input (if the naming method is Auto)
File : Text input (if the naming method is Arbitrary or Sequential number)
Name Type Auto, Arbitrary, Sequential number
File Type Binary (GBD), Text (CSV)
(2)-1 Sampling interval
This is used to set the sampling interval for data capture.
The table below shows the number of measuring channels*
1
and sampling interval values that can be set.
If data uctuate due to noise, set the sampling interval to a value which enables the digital lter function.
Number of Measuring
Channels*
1
Allowed Sampling Interval
Sampling Interval which enables
Digital Filter
1CH 10ms or slower *
2
50ms or slower
2CH 20ms or slower *
2
125ms or slower
3CH to 5CH 50ms or slower *
2
250ms or slower
6CH to 10CH 100ms or slower 500ms or slower
*1: "Number of Measuring Channels" is the number of channels in which input settings are NOT set to "OFF".
(CHs in sensor and wireless sensor connected through the wireless unit are not included.)
*2:
When the temperature setting is performed in 10, 20, or 50 ms sampling interval, the data capture cannot be performed.
When using the digital lter, the AC power frequency to be used must be set accurately. Please perform
the setting accurately in reference to “(6)-8 AC Line Frequency”.