3-23
CHAPTER 3 Settings and Measurement
(1)-3 Range
This is used to select the range of measurement.
Input item Description
DC 20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50, 100V 1-5 V
TEMP No selection (200C xed)
RH No selection (100% xed)
Available SPAN Settings
<Voltage Ranges>
Range Maximum SPAN Minimum SPAN Minimum Resolution
20mV
-
22.000 to +22.000mV 0.200mV 0.001mV
50mV
-
55.00 to +55.00mV 0.50mV 0.01mV
100mV
-
110.00 to +110.00mV 1.00mV 0.01mV
200mV
-
220.00 to +220.00mV 2.00mV 0.01mV
500mV
-
550.0 to +550.0mV 5.0mV 0.1mV
1V
-
1.1000 to +1.1000V 0.0100V 0.0001V
2V
-
2.2000 to +2.2000V 0.0200V 0.0001V
5V
-
5.500 to +5.500V 0.050V 0.001V
10V
-
11.000 to +11.000V 0.100V 0.001V
20V
-
22.000 to +22.000V 0.200V 0.001V
50V
-
55.00 to +55.00V 0.50V 0.01V
100V
-
110.00 to +110.00V 1.000V 0.01V
1-5V
-
5.500 to +5.500V 0.050V 0.001V
<Temperature Ranges>
Range Maximum SPAN
Minimum SPAN
(p-p)
Measurement
Range
Minimum
Resolution
K -270 to +2000℃ 50℃ -200 to +1370℃ 0.1
J -270 to +2000℃ 50℃ -200 to +1100℃
T -270 to +2000℃ 50℃ -200 to +400℃
R -270 to +2000℃ 50℃ 0 to +1600℃
E -270 to +2000℃ 50℃ -200 to +800℃
B -270 to +2000℃ 50℃ +400 to +1820℃
S -270 to +2000℃ 50℃ 0 to +1760℃
N -270 to +2000℃ 50℃ -200 to +1300℃
W -270 to +2000℃ 50℃ 0 to +2000℃
<Humidity Range>
Range Maximum SPAN Minimum SPAN (p-p) Minimum Resolution
100% 0 to +110% 1.0% 0.1%