3-28
CHAPTER 3 Settings and Measurement
(1)-7 Logic/Pulse
This is used to select the processing method for digital input.
Selection item Description
Off Digital input measurement is disabled.
Logic Digital input is processed as logic signals.
Pulse Digital input is processed as pulse signals.
(1)-8 Input
This is used to set the pulse measurement mode. This setting is available only if Pulse is selected in (1)-7.
Selection item Description
Off Pulse input measurement is disabled.
Revol. The number of pulses per sample interval is counted and converted to the number of revolutions per
minute.
Counts Captures the cumulative number of pulses for each sampling interval from the start of measurement.
Inst. Captures the number of pulses for each sampling interval.
(1)-9 Filter
This is used to set the lter for input.
Selection item Description
Off Disables hardware lter.
On Enables hardware lter. It is effective in a noisy environment.
The lter is approximately 30 Hz (-3 dB).
(1)-10 Pulse Slope
This is used to set the slope (direction) to count the number of pulses. This setting is available only if Pulse
is selected in (1)-7.
Selection item Description
H Counts the rising edges of pulses.
L Counts the falling edges of pulses.
•
Allocation of CH number of logic and pulse data is as shown below.
Bit 0 to 3: Main unit data
Bit 4 to 7: Wireless sensor data
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
MSB LSB