3-54
CHAPTER 3 Settings and Measurement
Dead zones of trigger and alarm levels
Trigger and alarm levels are provided with a dead zone in order to prevent false detection due to noise.
The following gure shows the dead zone.
<Mode: Rising>
Trigger (alarm) point
Dead zone level Dead zone level
<Voltage>
• Level setting value 1% of negative range
(1V range: Level setting value -0.01V)
<Temperature>
• Level setting value -0.3°C
Trigger (alarm) level
<Mode: Falling>
Trigger (alarm) point
<Mode: Rising> <Mode: Rising>
<Voltage>
• Level setting value 1% of positive range
(1V range: Level setting value +0.01V)
<Temperature>
• Level setting value +0.3°C
Trigger (alarm) level
The upper limit values within the mode range and the lower limit values out of the mode range are dead
zone level as well.
When the detected alarm drops below the dead zone level, it is cleared. (When the alarm is not retained)
<Example>When measuring temperature, the level released after the alarm generated is shown below.
• Rising setting: Setting value
-
0.3℃
• Falling setting: Setting value +0.C
In the case of pulse, dead zone is not provided.