19ENG - Version 2.2 - P/N 04131.02
Maintenance
Daily Check
1. Check the accessories/consumables for completeness.
2. Perform a visual check of the device and the mechanical
connections.
3. Switch on the compact device and check if the internal self-
test results in any error messages.
4. Select manual debrillation mode. Select an energy of 200 J
with the jog dial or the softkeys.
5. Release internal shock.
a) Paddles in holder: release with both paddle
buttons.
b) Therapy master cable in the base socket: re-
lease with the Shock key at the monitoring unit.
6. Print protocol by pressing the Browser key
and check the results for correctness.
7. Perform connection test by separating the
modules and connecting them again. Check connection
status in the status line.
Note
For further instructions for the daily check see chapter 9.2
and the checklist Functional Check in Appendix B of the User
Manual.