You can enter the Calibration Mode after accessing the Setup Mode to view and/or change
parameter settings (see Setup Mode on page 15).
To
access the Calibration Mode when viewing any setup parameter:
1. From the Setup Mode, press [ZERO].
DISPLAY READS Enter ~ =Cal!
2. Press [TARE] or [SAMPLE/Enter].
DISPLAY READS First ~ Zero? ~ -0.26
FAST CALIBRATION
Fast Calibration allows calibration of the Model 370/375 scale system without accessing the
Setup Mode.
To
access Fast Calibration from the Weigh Mode (370):
1. From the Weigh Mode, press [ZERO] + [SELECT].
DISPLAY READS Setup~Enter ~ Code!
2. Press [ZERO] [PRINT] [UNITS] [TARE].
DISPLAY READS Fast ~ Cal!~First ~ Zero? ~ -0.26
To
access Fast Calibration from the Weigh Mode (375):
1. From the Weigh Mode key in 100 [SELECT].
DISPLAY READS Setup~Enter ~ Code!
2. Key in 54321 [SAMPLE/Enter].
DISPLAY READS Fast ~ Cal!~First ~ Zero? ~ -0.26
Fast Calibration can also be accessed if the following data stream is received via the comm port:
100%s54321%e
PERFORMING CALIBRATION
Calibration always begins by establishing a zero (no-load) reference. A complete calibration also
requires establishing a span (test load) reference. This section details various methods for
obtaining zero and span references.
Establishing Zero
The Model 370/375 provides five methods for obtaining a zero (no load) calibration reference,
First Zero, Last Zero, False Zero, Only Zero, and Cal Reset.
Press [ZERO] or [CLR] during calibration to back up one step in the procedure.