66
7. Enter 2500.
DISPLAY READS 2500
8. Press [SAMPLE/Enter] to establish span.
DISPLAY READS 2500.
9. Pause for motion delay.
DISPLAY READS Cal ~ Good? ~ 2500.
10. Press [SAMPLE/Enter] to accept calibration.
DISPLAY READS Enter ~ =Stor
11. Press [SAMPLE/Enter] to save calibration.
DISPLAY READS Enter ~ =End
12. Press [SAMPLE/Enter] to exit calibration.
DISPLAY READS 5055.
13. Remove the calibration weight.
DISPLAY READS 00.
ONLY ZERO
Only Zero is used to establish a new calibration zero without affecting the span. This is useful for
correcting changes to the scale's dead load, for example adding safety rails to a truck scale
platform.
Only
Zero Calibration Example (370):
1. From the Weigh Mode, press [ZERO] + [SELECT].
DISPLAY READS Setup
2. Press [ZERO] [PRINT] [UNITS] [TARE].
DISPLAY READS Fast ~ Cal~First ~ Zero? ~2640.
3. Press [SELECT] [SELECT] [SELECT].
DISPLAY READS Only ~ Zero? ~ 2640.
4. Remove any load on the scale.
DISPLAY READS Only ~ Zero? ~ 2620.
5. Press [TARE] to establish zero.
DISPLAY READS 00.
6. Pause for motion delay.
DISPLAY READS Cal ~ Good? ~ 00.
7. Press [TARE] to accept calibration.
DISPLAY READS Enter ~ =Stor
8. Press [TARE] to save calibration.
DISPLAY READS Enter ~ =End
9. Press [TARE] to exit calibration.
DISPLAY READS 00.
Only
Zero Calibration Example (375):
1. From the Weigh Mode, press 100 [SELECT].
DISPLAY READS Setup
2. Press 54321 [SAMPLE/Enter].
DISPLAY READS Fast ~ Cal~First ~ Zero? ~2640.
3. Press [SELECT] [SELECT] [SELECT].