OPERATION 
157 
Please see the DMM chapter for details on the 
different measurement modes and connections. 
See from page 143.  
1.  From the EE Apps tray select the Advanced 
DMM icon. 
2.  From the measurement selection wheel select 
the measurement type. 
3.  The current measurement, minimum and 
maximum measurement will be shown as soon 
as the measurement type is selected.  
4.  Data is automatically logged as soon as a 
measurement function is selected. Logged data 
is plotted as a green line in the trend plot. 
5.  Press the Time button and select the logging 
time. Press OK to confirm the logging duration 
time. 
6.  The time is shown as the X-axis of the trend 
plot.  
40s, 3m, 6m, 15m, 30m, 1hr, 3hr, 6hr 
7.  To pause the measurement and data logging, 
press the Hold button. 
8.  To reset the logging, press the Reset button. 
To exit from the Advanced DMM app, press the 
Exit button. The advanced DMM will retain its 
state even after exiting, allowing data logging 
to continue.