GOM-802 User Manual
34
Temperature Compensation
Background
If the resistance of a DUT at a particular temperature is
needed the compensation function can be used. This
function can simulate the resistance of a DUT at a
desired temperature. If the ambient temperature and the
temperature coefficient of the DUT are known, it is
possible to determine the resistance of a DUT at any
temperature.
1. Select the
Temperature
compensation
mode.
Press SHIFT + COMP
T
EMP
to
access the main function mode.
Press the Up and Down arrow keys
to select the TC measurement mode
(shown above).
There are three different modes:
(resistance), (temperature)
and
(temperature compensation)
mode.
Press ENTER to confirm.
2. Temperature
compensation
mode appears.
Edit the desired temperature and the
temperature coefficient of the DUT.
Use the Left and Right arrow keys to
select a digit and use the Up and
Down arrow keys to edit the digit.
Desired
Temperature
0~99.9
Temperature
Coefficient
0000~9999
Press ENTER to confirm the
selection and start measurement.