APPENDIX 
  51 
Specifications 
 
Conditions 
Background 
The specifications are applicable under the following 
conditions: 
  A 1-year calibration cycle. 
  An operating temperature of 18 to 28 ˚C (64.4 to 
82.4˚F). 
  Relative humidity not exceeding 80%. 
  Accuracy is expressed as ±(percentage of reading + 
digits). 
  The instrument requires 30 minutes warm-up time to 
achieve rated accuracy. 
  The power cord protective grounding conductor must 
be connected to ground. 
 
Resistance Measurement 
 
30000 counts (speed : 7 readings/second) 
Range  Resolution 
Measuring 
Current  Accuracy 
Open-Terminal 
Voltage 
30mΩ 1μΩ 
1A approx.  0.1%+6* 
~1.772V 
300mΩ  10μΩ 
100mA  0.05%+6* 
~1.772V 
3Ω 100μΩ 
100mA 0.05%+3 
~1.772V 
30Ω  1mΩ 
10mA  0.05%+2 
~1.772V 
300Ω 10mΩ 
1mA 0.05%+2 
~1.772V 
3kΩ  100mΩ  100μA 
0.05%+2 
~1.770V 
30kΩ 1Ω 100μA 
0.05%+2 
~4.570V 
300kΩ  10Ω  10μA 
0.05%+2 
~4.526V 
3MΩ 100Ω 1μA 
0.05%+2 
~4.148V 
3000 counts Accuracy  (speed: 30 readings/second)  ±5 counts ** 
*When the instrument is set to 30mΩ or 300mΩ ranges, the resistance value 
will be changed while connecting or disconnecting the test lead to the panel 
due to the different temperature between internal and external parts of the 
instrument. Therefore, please must wait 1 minute in order to obtain an 
accurate value after the test leads have been connected or disconnected.   
* When Kelvin clips are used to resume testing after a long period of time, 
please wait for a short time to stabilize the measurement. 
** In high speed mode, first ensure that the instrument has been grounded 
through the power cord.