EasyManuals Logo

GW Instek GPP-3323 User Manual

GW Instek GPP-3323
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #124 background imageLoading...
Page #124 background image
GPP Series User Manual
124
Function
Query the specified measurement function in all
channels
Description
< function >: VOLTage,CURRent,POWEr
Example
:MEASure:VOLTage:ALL?
Reads back the all channels voltage value
Note
About CH3 of GPP-3323, Please refer to
" :MEASure[1|2|3|4] :<function>?"
Command
VOUT<X>?
Function
Returns the actual output voltage.
Description
X 1= CH1, 2= CH2, (4323: 3= CH3, 4=CH4)
Example
VOUT1?
Returns the CH1 output voltage
Note
About CH3 of GPP-3323, The query voltage will
return the set value.
Command
IOUT<X>?
Function
Returns the actual output current.
Description
X 1=CH1, 2=CH2, (4323: 3= CH3, 4= CH4)
Example
IOUT1?
Returns the CH1 output current
Note
About CH3 of GPP-3323, The query current will
return zero.