REMOTE CONTROL
133
Note
About CH3 of GPP-3323, Please refer to
":SOURce[1|2|3|4]:VOLTage < NRf >"
Command
:SOURce[1|2|3|4]:VOLTage?
Function
Queries the output voltage setting.
Example
:SOURce2:VOLTage?
Queries the output voltage setting on CH2.
Command
VSET<X>?
Function
Returns the output voltage setting.
Description
X 1= CH1, 2= CH2, (4323: 3 = CH3, 4= CH4)
Example
VSET1?
Returns the CH1 output voltage setting.
Command
:SOURce[1|2]:RESistor < NRf >
Function
Set the resistor value
Description
NRF
1-1000
Example
:SOURce2: RESistor 1000
Set the resistor value as 1000Ω for CH2
Command
:SOURce[1|2]:RESistor?
Function
Queries the setting resistor value.
Example
:SOURce2: RESistor?
Queries the setting resistor value and returns the
readback value for CH2.
Command
:SOURce:CURRent:ALL?

Table of Contents