Query
The F query returns the status of the power
supply.
F<CR>
ASCII
46 0D
HEX
F f1f2f3f4f5f6<CR><LF>
F
Status Byte character
f1f2f3f4f5f6
f1 :Relay Status:
0: off, 1: on
f2 :Temperature Status
0:normal, 1:overtemp
f3 : Step mode:
0: normal, 1:fine
f4 : Scroll wheel status:
0:locked, 1:unlocked
f5 : Remote Status:
0:normal, 1:remote
f6 : Lock Status:
0:unlocked, 1:locked
F<CR>
F101100<CR><LF>
Returns the status of the power
supply.
Relay on, temperature normal,
fine step mode, scroll wheel
unlocked, remote status normal,
panel unlocked.