SECTION 2, continued
17
2.2.2 Measurement Hints
Always cap the sample cell to prevent spillage of sample into the instrument.
Always close the sample compartment lid during measurement.
Install the appropriate Filter Assembly. The EPA Filter Assembly must be
installed for measurements requiring EPA reporting.
Do not leave a sample cell in the cell compartment for extended periods
of time.
Leave the instrument on 24 hours a day if the instrument is used regularly.
Empty the cell compartment and turn off the power if the instrument is stored
for extended periods of time.
Always use clean, scratch-free sample cells and caps.
Always apply silicone oil.
Always observe measurement techniques
9. Select the appropriate
measurement unit (NTU,
EBC or NEPH) by
pressing the
UNITS/EXIT
key.
10. Read and record the
results.
Note: A time- and date-
stamped measurement
record can be printed or
transmitted by pressing the
PRINT key.