EasyManua.ls Logo

Hach 2100AN - Diagnostic Codes

Hach 2100AN
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
54
SECTION 5, continued
5.2.2 Automated Flow-Cell Kit (Low Pressure)
The Automated Low Pressure Flow-Cell Kit (Cat. No. 47450-00, 115 Vac or
47450-02, 230 Vac), uses a Flow Valve Module for control of sample flow
(see Figure 11). The kit contains a remote control cable used with the Model
2100AN Turbidimeter for automated operation. Refer to Section 5.2.1.1 and
Section 5.2.1.2 for assembly instructions. Omit step 5 in Section 5.2.1.1;the
Collection Drain Assembly is not provided with the Automated Flow-Cell Kit.
Figure 11 Automated Low Pressure Flow Cell
5.2.2.1 Connecting Inlet and Outlet Tubing
1. Cut a 53-cm (21") piece of clear,
1
/
8
" I.D. Tygon tubing. Install it between the
Flow-Cell Inlet Reservoir and Flow Valve Module inlet.
Note: Use the tubing supplied with the kit (or its equivalent). Tubing lengths are
approximate. Avoid using excess tubing because it causes air locking, and delays
measurement response time.
2. Cut a 31-cm (12") piece of clear,
1
/
8
" I.D. Tygon tubing. Install it between the
Flow Valve Module outlet and the Flow-Cell inlet.
3. Cut two 25-cm (10") pieces of clear,
1
/
8
" I.D. Tygon tubing. Install them
between the top and bottom Flow Cell drain fittings and the “Y” connector.
4. Cut a 11-cm (4") piece of clear,
1
/
8
" I.D. Tygon tubing. Connect one end to
the remaining Y” connector. Pass the tubing under the support stand base
as illustrated in Figure 11. Install the
1
/
8
"x
1
/
4
" reducer on the other end of the
tubing.
5. Cut a 50-cm (20") piece of clear,
1
/
4
" I.D. Tygon tubing for the drain line.
Connect one end to the
1
/
8
"x
1
/
4
" reducer, and run the other end to a
suitable drain.

Table of Contents

Other manuals for Hach 2100AN

Related product manuals