66
SECTION 6, continued
6.1.2 Setting Recorder Maximum Value
This function is used to set the maximum value of the recorder output for the
current units (selected measurement mode).
1. Enter the setup mode by pressing
SETUP. The mode display flashes.
2. Select the number
15 using the edit keys followed by pressing the ENTER key.
The decimal point will flash.
3. Move the decimal point to the desired location using the right arrow key, and
then press
ENTER to accept the new location. The left digit will flash.
4. Set the desired recorder maximum value using the edit keys. Press
ENTER to
accept the new setting. Pressing
UNITS/EXIT at any time, prior to accepting the
new value, exits the setup mode leaving original values intact (see note under
Section 6.1.1).
6.1.3 Setting Recorder Minimum Output
This function moves the recorder minimum output in a positive or negative
direction to calibrate the recorder.
1. Enter the setup mode by pressing
SETUP. The mode display flashes.
2. Select the number
16 using the edit keys. Press ENTER. The display will show
UP/DN RTN.
3. Press the up arrow key to increase the minimum output in a positive direction.
Press the down arrow key to decrease the minimum output in a negative
direction. Press
ENTER to accept the new setting.
4. Press
SETUP to exit the setup mode. Pressing UNITS/EXIT at any time, prior to
accepting the new value, exits the setup mode leaving original values intact.
6.1.4 Setting Recorder Full-Scale Output
This function moves the recorder full-scale output in a positive or negative
direction to calibrate the recorder.
1. Enter the setup mode by pressing
SETUP. The mode display will flash.
2. Select the number
17 using the edit keys. Press ENTER. The display will show
UP/DN RTX.
3. Press the up arrow key to increase the full-scale output in a positive direction.
Press the down arrow key to decrease the full-scale output in a negative
direction. Press
ENTER to accept the new setting.
4. Press
SETUP to exit the setup mode. Pressing UNITS/EXIT at any time, prior to
accepting the new value, exits the setup mode leaving original values intact.